Contact an Account ManagerAnalysis of antibiotics and wipe samples from environmental and biological monitoring
Summary
Type: HPLC
Application #: 118273
Catalog#: 760001.20 - MN HPLC Column, Nucleodur C18 ec, 100A, 5um, 2x125mm
Detection: LC-MS/MS (triple quadrupole, electrospray, positive and negative mode, 5000 and -4500 V, settling time 700 ms)
Conditions: Eluents: 0.1% formic acid in water (phase A) and 0.1% formic acid in acetonitrile (phase B), (v/v)
Gradient: 95% A for 1 min, 95 to 30% A in 14 min, 30% A for 2 min, 30 to 95% A in 1 min, 95% A for 6 min
Flow rate: 0.3 ml/min
Temperature: 30 °C
Authors: Türk J Reinders M Kiffmeyer T K Schmidt K G Schöppe G Kuss H M Dreyer D
Source: Institute of Energy and Environmental Technology G